Chief & Presenting Author: Dr. Ram Lolarakh
Co Author(s): Prof.Aparajita Chaudhary, Dr.Jagriti Rana
Abstract
To compare central corneal thickness measurements obtained by Pentacam with those obtained by IOL Master 700, Anterior segment optical coherence tomography and Specular microscopy in normal healthy eyes.The mean CCT [± standard deviation (SD)] for Pentacam, IOL Master 700, Cirrus AS-OCT and Tomey Specular microscopy were Pentacam (Oculus), AS-OCT (Cirrus), IOL Master 700 and Specular microscopy (Tomey) were 523.75 (±27.75), 525.29 (±28.81),517.13 (±28.43) and 512.82 (±27.60) μm, respectively. All the means were significantly different from one another (P < 0.000). The differences between pairs of mean central corneal thickness (CCT) for Pentacam and IOL Master, Pentacam and anterior segment- optical coherence tomography (AS-OCT), and Pentacam and Specular microscopy are statistically significant.